2. About broncolor Radio System
8.
Press the Measuring Button 6.
The ash will re, and the ash duration time and the measured value (F-stop) for
input ISO sensitivity and shutter speed will be displayed.
NOTICE
The ash duration time and graph are displayed in the Flash Duration Analysis Radio
Triggering Mode, however, they cannot be stored in the memory. They are cleared if the
Measuring Mode is changed or the POWER switch is turned OFF.
The incident light measurement can only be used in Flash Duration Analysis Radio
Triggering mode.
In case of the followings, please follow “Flash Duration Analysis Cord Mode”( ).
When ring the ash, if the ash brightness is lower than the ambient light, the meter
may fail to detect the light.
Rapid start uorescent lamps and special lighting are sometimes mistaken for ash and
accidentally measured.
Even if the ash does not re, when a sudden light change occurs in the light receptor,
measurement may be made.
The waveform of a ash bulb has a slight slope and there is a possibility that the light
meter cannot recognize the ash bulb.
If the measured ash duration time is longer than the input
shutter speed, an appropriate F-stop cannot be measured.
The yellow “Under” indication appears. In this case, slower
the shutter speed than the ash duration time and measure
again.
[Measuring Screen]
Measured Value (F-stop)
Flash Duration Time
[Measuring Screen]