21
2. About broncolor Radio System
7.
Press the Measuring Button 6.
The measured value (F-stop) will be displayed. Press the Measuring Button
6
again to re all ashes in the same Studio Address and measure until the desired
F-stop is displayed.
The accumulated measured value (F-stop) and the cumulative count will be
displayed.
If you want to adjust the output power of each ash separately, go to Flash Power
Control Screen to make a measurement and adjustment.
In the Flash Power
Control Screen, the Multi (cumulative) ash measurement is not
available. (
P14)
NOTICE
In case of the following, please follow “Cord Multi (Cumu.) Flash Mode” ( ).
When ring the ash, if the ash brightness is lower than the ambient light, the meter
may fail to detect the light.
Rapid start uorescent lamps and special lighting are sometimes mistaken for ash
and accidentally measured.
Even if the ash does not re, when a sudden light change occurs in the light receptor,
measurement may be made.
The waveform of a ash bulb has a slight slope and there is a possibility that the light
meter cannot recognize the ash bulb.
The EV scale cannot be displayed in the Radio Triggering Multi (Cumu.) Flash mode.
NOTE
With broncolor radio system, individual ash triggering is not available for each Lamp
Address. All ashes in the same Studio Address will re even if a particular Lamp
Address is selected. For measuring the ash individually, turn off the power of ash units
not in use. Or use the retracted lumisphere of L-858D to measure the light in front while
all ashes re.
For red-eye prevention and auto-ash light adjustment, some devices can pre-ash
before main ash burst. With the normal setting, the light meter will measure the pre-
ash bursts and not the main ash burst. To take a successful reading, activate the pre-
ash feature in the Tool Box. Please refer to the L-858D Operating Manual.
For radio frequencies, refer to “5. Radio Studio Address Frequencies”. (
P36)
[Measuring Screen]
Measured Value
(F-stop)
Cumulative Count