20 Set the measurement parameters
8. Set the measurement parameters
"Measurement settings" menu allows to set units, temperature compensation, pressure type and atmospheric
pressure.
To access this menu:
• From the main menu, go to "Measurement settings" and press OK.
10:09 Menu
Measurements / Tests
Measurement settings
Wireless connection
General settings
Fault tree analysis
Data recording
Device information
>
>
>
>
>
>
>
10:09 Measurement settings
Units
Temperature comp.
Pressure type
Atmo pressure
>
>
Relative
1013 mbar
8.1 Set the units
The following units can be set for these different parameters:
• Temperature: °C, °F
• Pressure: bar, psi, MPa, kPa
• Vacuum pressure: micron, Pa, hPa, mbar, mTorr, mmHg, inH
2
O, inHg, Torr
• Atmospheric pressure: mbar, hPa, inHg, bar, psi
• Weight: kg, lb, oz
"Measurement settings" menu is displayed.
• Press OK on "Units" line.
• Press OK on the required parameter line.
The list of available units is displayed.
• Select the unit with with ▲ and ▼ and press OK.
8.2 Set the temperature compensation
A temperature compensation factor has been set in the instrument to reduce the errors when measuring pipe
temperature in the main eld of applications. This reduces measuring errors for Si-RT2, Si-RT5 and Si-RT7 when
measuring surface temperature.
For measuring contact temperature, the compensation must be turned on and for measuring surface temperature,
the compensation must be turned off.
When doing calibration with temperature sensor in a bath the temperature compensation should be turned off.
"Measurement settings" menu is displayed.
• Press OK on "Temperature comp." line.
• Press OK on "Temperature compensation" line to switch between ON and OFF.
• Select the type of probe to be used for the temperature compensation:
• Press OK on "Wired Port 1" or "Wired Port 2".
• Press OK on "Velcro" or "Clamp".
8.3 Set the pressure type
"Measurement settings" menu is displayed.
• Press OK on "Pressure type" line to switch between "Relative" and "Absolute".
8.4 Set the atmospheric pressure
"Measurement settings" menu is displayed.
• Press OK on "Atmo pressure" line.
• Dene the atmospheric pressure value with + and - and press OK.
The atmospheric pressure value must be between 400 and 1200 mbar.