16 Connect a probe and the scale
6. Connect a probe and the scale
6.1 Wired probe
The device has two inputs allowing the connection of wired temperature probes:
• Si-RT2 temperature clamp
• Si-RT5 temperature probe with hook-and-loop fastener
6.2 Wireless probe
The following wireless probes are avalaible:
• Si-RT7 temperature clamp
• Si-TH4 temperature and humidity (psychrometer) probe
• Si-RP4 pressure probe
• Si-RV4 vacuum probe
When using a wireless probe for the rst time, you must scan the probe before using it.
6.2.1 Scan a probe
The main menu is displayed.
• Turn on the wireless probe.
LED turns on and is xed.
• Make sure wireless communication is activated.
• Press OK on "Wireless connection" line.
• Press OK on "Scan devices" line.
The probe name (e.g. Si-RT7) and its serial number are displayed.
• Press OK.
The connection is in progress then the probe is connected:
• the LED on the probe blinks.
• probe name, type of probe, serial number, address and rmware version of the probe are dis-
played.
• Press OK.
• For Si-RT7 probes only: select the usage for this probe: press OK on T1, T2 or T3.
• For Si-TH4 probes only: select the usage for this probe: press OK on WB or DB temperatures.
• Press ESC to back to main menu.
Once a probe is paired to the manifold, you can perform your measurements.
You can turn off the probe then turn it on. It will be recognized by the manifold, there's no need to scan it
again.