5. F 200: Configuration of the channels and measurement units
This folder allows to activate the channels and to set measurement unit for each channel.
5.1. Activate a channel
The transmitter is powered on.
Press OK.
Enter the activation code (see page 9).
Press OK.
Press Up key to go to F 200 folder.
Press twice OK.
“SDE” is displayed below “F 200”.
Press OK.
“SDE” for probe or “OFF” for deactivated channel or “PRES” for pressure blinks.
Press Up and Down key to select:
“SDE”: channel is activated with a measurement probe
“PRES”: channel is activated with a pressure board
“OFF”: channel is deactivated
Press OK.
The last zero of F200 blinks.
“F200” folder allows to activate the channel 1 of the transmitter.
To activate channels 2 and 3 go to the following folders:
F210 for channel 2
F220 for channel 3
Perform the same procedure as for the channel 1.
5.2. Assign a measurement unit to a channel
The transmitter is powered on and a channel is activated.
“F 200” folder is displayed.
Press Up key.
“F 201” sub-folder is displayed with the unit corresponding to the channel 1 below.
Press OK.
The unit blinks.
Press Up and Down keys to select the required unit.
Press OK.
“---” sign means that the channel is deactivated.
“F201” sub-folder allows to select the unit for the channel 1.
To select the unit for channels 2 and 3, go to the following folders:
F211 for channel 2
F221 for channel 3
Perform the same procedure as for the unit selection for the channel 1.
F 200: Conguration of the channels and measurement units
13

Bekijk gratis de handleiding van Sauermann CPE 311-S, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkSauermann
ModelCPE 311-S
CategorieNiet gecategoriseerd
TaalNederlands
Grootte3055 MB