Note
Do not clean the mesh or the accessories mechanically us-
ing a brush or similar device, as this could cause irreparable
damage and it will mean that the best treatment results can
no longer be guaranteed.
Please consult your doctor about the additional requirements
in terms of the hygienic preparation needed (hand care, han-
dling of medication/inhalation solutions) for high-risk groups
(e.g. patients with cystic fibrosis).
Cleaning
The mesh atomiser (2) and the accessories used such as the
mouthpiece, mask, etc. must be washed with warm water of
max. 40°C for approx. 5 minutes after each use. Dry the parts
carefully using a soft cloth. Put the parts together again when
they are completely dry and place them in a dry, sealed con-
tainer or disinfect them.
Remove the mouthpiece and the relevant accessory from the
mesh atomiser (2).
Pour 6ml of clean water into the medication container (1) and
switch the device on to first clean the mesh of any medication
residue (please nebulise all of the liquid).
Disassembling
First remove the cover of the mesh
atomiser (1) by turning it anti-clock-
wise.