- 6 -
only available when the number of active channels is no more than 8),
channel label and so on.
Note: When the probe is connected to the oscilloscope for the first
time or the temperature change is more than 5 degrees, you are
recommended to calibrate the probe zero using the calibration
function in the LA menu (press LA Threshold Calibration) and
please disconnect all the connections to the PLA2216 input terminal
during the calibration.
4 Function Check: after finishing the above operations, the signal
under test will be displayed on the corresponding digital channel on
the oscilloscope screen. If no signal is displayed, please adjust the
oscilloscope to select proper general settings (such as the trigger
source, trigger level and trigger mode). If signal is still not displayed,
please check the electric connection and parameter settings again or
please try to use other probe (such as analog probe) to check the
signal state of the test point.