7.7 XY Mode
By default, this series oscilloscope uses the YT mode for waveform display window. In
YT mode, Y-axis indicates the Voltage and X-axis indicates the Time. Besides, it
supports XY display window. In this display window, X-axis and Y-axis indicate
voltage. The two input channels display from "Voltage-Time" to "Voltage-Voltage".
Enable the XY Mode
You can enable the XY display mode in the following ways.
• Click or tap the Windows button in the function navigation menu or on the
toolbar to enter the
Add Window menu. In the "Diagram" item, click or tap XY >
Add to enable the XY display mode.
• Click or tap the XY button in the function navigation menu or on the toolbar to
enable the XY display mode.
• In the "Horizontal" menu, check the checkbox of XY to enable the XY mode.
Configure the XY Mode
Click or tap at the upper-right corner of the XY display window to enter the XY
configuration menu.
Figure 7.3 XY Menu
• Source: Click or tap the drop-down button of Source X to select the source
channel of the X-axis in the XY window. Click or tap the drop-down button of
Source Y to select the source channel of the Y-axis in the XY window.
• Grid: Please refer to
To Set the Screen Grid
to set the grid of the XY window.
Phase Deviation Measurement
In this mode, you can use the Lissajous method to measure the phase deviation of the
two input signals whose frequencies are the same. The following figure shows the
measurement schematic diagram of phase deviation.
Acquisition System
MHO2000 User Guide
68
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.