In Vertical menu, click or tap the Up/Down arrow icon at the right side of the
input field of Offset to increase or decrease the offset value or use the specified
multifunction knob to set the value. You can also click or tap the input field of
Offset to input a specific value with the pop-up numeric keypad.
Input Field
Decrease
Increase
5.4 Channel Coupling
You can remove unwanted signals by setting the coupling mode. For example, the
signal under test is a square waveform with DC offset.
Click or tap the channel status label at the bottom of the screen, and then the
Vertical menu is displayed. Click or tap the drop-down button of Coupling to select
the coupling mode.
When the coupling mode is "DC", both the DC and AC components of the signal
under test can pass the channel.
When the coupling mode is "AC", the DC components of the signal under test
are blocked.
When the coupling mode is "GND", the DC and AC components of the signal
under test are blocked.
After a coupling mode is selected, it is indicated in the channel status label at the
bottom of the screen, as shown in the figure below.
DC
AC
GND
TIP
When the input impedance is set to "50 Ω", the channel coupling is set to DC coupling by
force, and the Coupling menu is grayed out and cannot be modified.
To Set the Vertical System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. MHO2000 User Guide
47