Source Setting
• Source A
Click or tap the drop-down button of SourceA to select the analog channel or
digital channel. The current trigger source is displayed in the trigger information
label at the top of the screen.
Only when we select the channel (that has been input with signals) as the trigger
source, can we obtain a stable trigger.
• Edge A
Click or tap the trigger edge type ("Rising" or "Falling") of Source A in Delay
trigger under the
EdgeA menu.
• Source B
Click or tap the drop-down button of SourceB to select the analog channel or
digital channel. The current trigger source is displayed in the trigger information
label at the top of the screen.
Only when we select the channel (that has been input with signals) as the trigger
source, can we obtain a stable trigger.
• Edge B
Click or tap the trigger edge type ("Rising" or "Falling") of Source B in Delay
trigger under the
EdgeB menu.
Set the Trigger Condition
Sets the time limit condition of Delay trigger for When menu item.
• >: triggers when the time difference (∆T) between the specified edges of Source
A and Source B is greater than the preset time lower limit. Click or tap the input
field of
Lower to set the delay time lower limit with the pop-up numeric keypad.
You can also use the specified knob to set it.
• <: triggers when the time difference (∆T) between the specified edges of Source
A and Source B is smaller than the preset time upper limit. Click or tap the input
field of
Upper to set the delay time upper limit with the pop-up numeric keypad.
You can also use the specified knob to set it.
• < >: triggers when the time difference (∆T) between the specified edges of
Source A and Source B is greater than the lower limit of the preset time and
smaller than the upper limit of the preset time. Click or tap input field of
Upper
and
Lower respectively to set the delay time upper limit and lower limit with the
pop-up numeric keypad. You can also use the specified knob to set them
respectively. The lower time limit must be smaller than the upper time limit.
• > <: triggers when the time difference (∆T) between the specified edges of
Source A and Source B is smaller than the lower limit of the preset time or
To Trigger the Oscilloscope