Trigger Position
After selecting the window type, click or tap Position to further specify the time point
of trigger by selecting the trigger position.
Enter: triggers when input signal enters the specified trigger level range.
Exit: triggers when input signal exits the specified trigger level range.
Time: triggers when the accumulated hold time since the input signal entered
the specified trigger level range is equal to the window time. When this type is
selected, you can set the window time. Its available range is from 1 ns to 10 s.
Set the Trigger Level
After the setting is completed, you need to adjust the trigger level to correctly trigger
the signal and obtain a stable waveform.
Level A: only adjusts the upper limit of the trigger level, and the lower limit of
the trigger level remains unchanged.
Level B: only adjusts the lower limit of the trigger level, and the upper limit of
the trigger level remains unchanged.
Linkage: When you check the checkbox of
Linkage, the upper and lower trigger
level can be adjusted synchronously. The trigger level deviation (the difference
between the upper limit and lower limit of the trigger level) remains unchanged.
You can set "Level A" and "Level B" with the pop-up numeric keypad. You can also
rotate the Trigger Level knob on the front panel or the specified multifunction knob
indicated in the input field of
Level A and Level B to set the value. For details, refer to
descriptions in
Trigger Level
.
TIP
Press down the trigger level knob and rotate it to adjust Level A, Level B, and Level AB (when
linkage is enabled) in sequence.
8.7.10 Delay Trigger
In Delay trigger, you need to set Source A and Source B. The oscilloscope triggers
when the time difference (∆T) between the specified edges (Edge A and Edge B) of
Source A and Source B meets the preset time limit, as shown in the figure below.
Edge A and Edge B must be two neighboring edges.
To Trigger the Oscilloscope
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved. MHO2000 User Guide
95