RaySafe X2 VOLT
VOLT
MEASURE WITH VOLT SENSOR
Connect the Volt sensor to the device using the Volt cable.
A live value is displayed in the lower left corner of the screen. The live value is the average volt-
age, recalculated and updated four times per second.
Note! To ensure functionality and safety, only use Volt cables provided by RaySafe.
To get other measurement parameters, in addition to voltage (V) and time (s), select an appli-
cable conversion factor. Swipe right from the home screen to access the quick settings with the
conversion factors.
A conversion to kV will add a calculated tube voltage (kV) value to the measurement.
A conversion to mA will add calculated tube current (mA) and tube charge (mAs) values to the
measurement.
Press the home button to go back to the home screen.