RaySafe X2 MAM
Mount the X2 MAM sensor in the holder.
Place the holder centered on the table, tight to the chest wall.
Settings:
Philips MicroDose: Stop delay: 2 s. W/Al Philips, if kVp is desired.
Fischer SenoScan: Stop delay: 2 s. W/Al, if kVp is desired.
Adani: Stop delay: 2 s. W/Al, if kVp is desired.
Always use the compression paddle, positioned as high as possible. Do not use the AEC function-
ality.
MAM SENSOR: MEASUREMENT PARAMETER DEFINITIONS
Dose and HVL are calculated from all recorded data.
Dose rate is the average dose rate, calculated as dose/time.
Time starts the first time the dose rate waveform reaches 50% of peak, and ends the last time it
drops below 50%. Live readings are time since start trig.
kVp is calculated from an average of samples above 33 % of peak signal level.
Pulses are counted at every start trig with more than 4 ms since last end trig.
Pulse rate and dose per pulse are averages from the last 6 pulses.
Live readings of dose rate, kVp and HVL are moving averages. Final readings are based on the
whole measurement.