RaySafe X2 – MAM
MAM
MEASURE WITH MAM SENSOR
The MAM sensor is ready to measure dose and HVL for all anode/filter combinations without
changing any settings. Place the connected sensor centered in the field, with the crosshair 6 cm
(4 cm in the US) from the examination table front edge (A). The angle in the horizontal plane
has neglectable impact on the measurement result.
For kVp measurements, select your anode/filter combination in the sensor settings, or by swiping
right from the home screen to access the quick settings.
Note! When measuring kV on Mo/Rh, use the included 2 mm Al filter. Follow the instructions
printed on the filter.
Expose.
24