Pro-Digi User Manual – Test evaluation
CENTRAL BEAM ALIGNMENT
Use the center cone to determine if the X-ray beam is perpendicular to the image receptor and centered with
respect to the light field.
When the dot is inside the circle, the deviation from a perpendicular beam is less than 1.5°.
DYNAMIC RANGE
EXAMPLE All seven dierent shades of gray should be visible.
SPATIAL RESOLUTION
Use the zoom tool on a workstation to determine the line pairs per millimeter resolution. You should be able to
distinguish three separate lines.
EXAMPLE The high contrast resolution should not change over time and should be within the
values used in the resolution lead bar.
CONTRAST RESOLUTION
EXAMPLE All six low contrast objects should be visible and should not change over time.