AFTER EACH USE
Follow your local guidelines and medical protocol to download incident data from the FR3 data
card if one is installed. To ensure that data is not lost due to a full data card, Philips recommends
that you replace a used FR3 data card with a blank FR3 data card after each use.
Check the outside of the FR3 for signs of damage, dirt, or contamination. If you see signs of
damage, contact Philips for technical support. If the defibrillator is dirty or contaminated, clean it
as described below.
Replace any used, damaged or expired items.
Run a user-initiated test as described below.
Press the On/Off button to display the status screen then let the FR3 go into standby to be ready
for use. Note any messages displayed on the status screen prior to shutdown, and address them as
appropriate.
Return the FR3 to its storage location to be ready for use when needed.
USER-INITIATED TEST
To run the user-initiated test, follow the steps below. However, as long as the green Ready light is
flashing and the FR3 is not chirping, it is not necessary to test the defibrillator by initiating a test. This
uses battery power and drains the battery prematurely.
1. Press the green On/Off button to start the FR3. When voice prompts begin, press the button
again to display the status screen.
2. Before the progress bar on the status screen is filled, press the ADMINISTRATION option button.
3. On the ADMINISTRATION screen, press the
option button to scroll to USER-INITIATED
TEST, then press the
option button to select it. The FR3 gives a tone and begins the test.
4. Follow the instructions on the screen. The first part of the test is interactive. Follow the prompts
to press each button when instructed, or the test will fail. Then listen for two audio tones. If you
do not hear the tones, there is a problem with the speaker or the beeper. Contact Philips at
www.philips.com/AEDSupport for technical support.
5. The second part of the test is automatic. During the automatic part of the test, the screen displays
a progress bar indicating the test’s progress.
6. When the test is complete, the FR3 displays the results of the test, the device serial number, and
the software version, then displays the ADMINISTRATION screen again.
7. Press the On/Off button to display the status screen and put the defibrillator into standby.
8. If the FR3 or any attached accessory fails the self-test for any reason, the screen displays
information about the failure. If the FR3 passes the test, the green Ready light flashes to show that
the FR3 is ready for use.
Press ON for Patient Use
Administration OFF
Battery
Data Card Good
Good
Shutting Down
Wireless Data Transfer
System Information
Incident Data
ADMINISTRATION
Setup
User-Initiated Test
#:##:##
##
Press ON for Patient Use
Press Shock Button
to Run User-Initiated Test
7HVW3DVVHG
$GPLQLVWUDWLRQ 2))
6HULDO1XPEHU
6RIWZDUH9HUVLRQ
&;
35
8VHU,QLWLDWHG7HVW5HVXOWV
5HDG\IRU8VH