‘NORMAL MODE’
When the power supply is connected, the instrument:
activates the conguration digits’, and displays the code for the actual
input signal range (see section 5)
activates the decimal point (ash), showing the actual output signal
range (see section 8.3)
the instrument is in ‘normal mode’ of operation
HOW TO ACCESS THE ‘CONFIGURATION MODE’ AND THE ‘TOOLS MENU
With the instrument in ‘normal mode’ of operation, there is access to the
conguration mode’ and the ‘tools’ menu.
to access the ‘conguration mode’, press the SQ’ (<) key for 1 second
(see section 8.2)
to access the ‘tools’ menu, press the ‘UP’ (5) key for 1 second (see
section 8.4)
‘ECOFUNCTION (‘CONFIGURATION DIGITS’ ARE AUTOMATICALLY
POWERED OFF)
If there is no interaction from the operator for 60 seconds, the instrument
powers off the ‘conguration digits’. The decimal point remains active
(ashing), indicating that the instrument is working correctly.
HOW TO POWER ON THE ‘CONFIGURATION DIGITS’
To power on the conguration digits’, press one of the front keys SQ
(<) or ‘UP’ (5). This will power on the conguration digits’, and activate
the ‘normal mode’ of operation.
8.  Conguration system
CONFIGURATION SYSTEM
The isolated signal converters provide a fast and easy conguration
system, based on function codes. Conguration is done through an
internal key pad, accessible behind the front cover.
‘CONFIGURATION DIGITS’
The instrument provides two numerical leds, to inform about the actual
state of the instrument, active input and output signal ranges, and error
codes (see section 11).
CONFIGURATION KEYS
The conguration keys are located inside the instrument, behind the
removable front cover. To access the conguration keys, remove the
output signal terminal, and open the front cover (see section 7). The
instrument allows to block the key functions, by activating an access
code (‘password’ function) (see section 8.5).
8.1.  ‘Normal mode’ of operation
HOW TO ACCESS THE ‘CONFIGURATION MODE’
With the instrument in ‘normal mode of operation (see section 8.1),
press for 1 second the SQ’ (<) key. The horizontal leds lights from
bottom to top. When the upper led lights, the instrument activates the
conguration mode’.
If the key is released before activating the ‘conguration mode’,
the horizontal leds light downwards from top to bottom, and
the instrument returns to ‘normal mode’ of operation.
Inside the ‘conguration mode’ the decimal point is not active.
Inside the ‘conguration mode’, the ‘conguration digits’ display
the code of the active input signal range (see section 5). Press the ‘UP
(5) key to increase the code.
press once to increase the value in +1
maintain the key pressed to automatically increase the value
When the desired code is reached, press the ‘SQ’ (<) key. The leds light
a round shape while new conguration is stored, and the instrument
returns to ‘normal mode’ of operation.
When exiting the conguration mode without changes (‘rollback’
activation, selection of ‘inactive codes’, etc), the horizontal leds
light down from top to bottom, and the instrument returns to
‘normal mode’ of operation.
‘ROLLBACK’ FUNCTION
If there is no interaction from the operator for 30 seconds, the instrument
exits the ‘conguration mode’ and returns to ‘normal mode’ of operation,
discarding changes.
OUTPUT 4/20 mA AND 0/10 Vdc
Instruments can be congured for 4/20 mA or 0/10 Vdc output signal.
The congured output signal is identied with the active decimal point
at the ‘conguration digits’.
active decimal point to the right, for 4/20 mA output
active decimal point to the left, for 0/10 Vdc output
Table 4 | Decimal point indicates the active output signal
Output 0/10 Vdc
(decimal point to the left)
Output 4/20 mA
(decimal point to the right)
UP SQ UP SQ
8.2.  ‘Conguration mode
8.3.  Output 4/20 mA and 0/10 Vdc
HOW TO CONFIGURE THE OUTPUT SIGNAL TO 4/20mA AND 0/10 Vdc
With the instrument in ‘normal mode’ of operation, press the SQ’ (<)
key and the UP’ (5) key at the same time. The horizontal leds lights
from bottom to top. When the upper led lights, the actual decimal point
position is displayed. After 1 second, the new decimal point position
activates. Release the two keys, and the leds light a round shape while
the new conguration is stored, and the instrument returns to ‘normal
mode’ of operation.
Connect the output terminals according to the output signal