MAINTENANCE
To keep the device in good condition, it is recommended to use the
device at least once a month.
CALIBRATION
e device is factory calibrated.
e accuracy of the measurement can change aer 6 to 12 months
of normal use, depending on the number of tests performed and the
conditions of use.
If the device provides inconsistent test results, no results, or exceptionally
high or low results, it should be recalibrated by a qualied specialist.
Contact service for calibration.
If the total number of tests performed is 300 or more, or you have used
the device for more than a year since the last calibration (whichever
comes rst), it is recommended to send the device to a local distributor
for recalibration.