Factory reset to Default settings
10. Factory reset to Default settings
10.1 Factory default settings
SETUP - change test pass limits, test times and test lead resistance. SETUP is “test group based” as the PASS limit for a Class 1
insulation test is different to a Class 2 insulation test.
10.2 Factory Default Test Limits
Variant Model
Rpe, Rcont (Ω)
Rpe, Rcont (Ω) for
Ext lead
Rpe, Rcont (Ω) for
RCD
Class 1
Riso, Rins
(MΩ)
Class 2
Riso, Rins
(MΩ)
Ext lead
Riso, Rins
(MΩ)
Class 1 I
ea,
ileak(sub), ipe, ileak
(mA)
It, I
b
Class 2 iea,
I
t
(sub) (mA)
1xIΔN30 (ms)
5xIΔN30 (ms)
1xIΔN10 (ms)
5xIΔN10 (ms)
PAT250SX-UK 0.2 0.2 0.2 1 2 1 3.5 0.25 200 40 200 40
PAT250SX-DE,
PAT250SX-CH,
PAT250SX-EU
0.3 0.3 0.3 1 2 1 3.5 0.5 300 300
PAT250SX-AU 1 1 1 1 1 1 5 1 300 40