www.megger.com PAT250SX
29
SETUP
7. SETUP
Changing PASS limits and test times
1.
2. To select a TEST
GROUP to be
modified press the
relevant button:
or
or
3. Screen displayed 4. Keep pressing the
TEST GROUP button
to select the test to
be changed
Pass Limit
Default Rcont 0.01 Ω
1st press Rins 1.00 MΩ
2nd press l leak 3.50 mA
Test Time
3rd press Rcont 5: S
4th press Rins 5: S
5th press l leak 5: S
5. Example changing
Insulation pass limit
6. Pressing SETUP
button changes the
value
Note : Pressing QT
changes the direction
Default 1.00 MΩ
1st press 2.00 MΩ
2nd press 0.01 MΩ
3rd press 0.05 MΩ
4th press 0.25 MΩ
5th press 0.30 MΩ
6th press 0.50 MΩ
7. Example: R
ins
change to 2.00 MΩ
8. To SAVE changes to
setup
9. or, to edit new test
groups
10. When changes are
complete press the
Power button