Guidance and manufacturer’s declaration - electromagnetic Immunity
Radiated RF
IEC61000-4-3
(Test specifications for
ENCLOSURE PORT
IMMUNITY to
RF wireless communi
-
cations equipment)
Test
Frequency
(MHz)
Band
(MHz)
Service Modulation Modulation
(W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385 380 -
390
TETRA 400 Pulse
modulation
18 Hz
1.8 0.3 27
450 430 -
470
GMRS 460,
FRS 460
FM
± 5 kHz devi
-
ation
1 kHz sine
2 0.3 28
710 704 -
787
LTE Band
13,
17
Pulse
modulation
217 Hz
0.2 0.3 9
745
780
810 800 -
960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
18 Hz
2 0.3 28
870
930
1720 1700 -
1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
217 Hz
2 0.3 28
1845
1970
2450 2400 -
2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2 0.3 28
5240 5100 -
5800
WLAN 802.11
a/n
Pulse
modulation
217 Hz
0.2 0.3 9
5500
5785