EMC TABLES
Guidance and manufacturer’s declaration of electromagnetic immunity
The Mebby NOSE CLEAN Nasal Aspirator is intended for use in the
electromagnetic environment specied below. The customer or the
user of the Mebby NOSE CLEAN Nasal Aspirator should assure that it
is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electroma-
gnetic
environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8kV contact
±15kV contact
±8kV contact
±15kV contact
Floors should
be wood, con-
crete or ceramic
tile.
If oors are co-
vered with syn-
thetic material,
the relative hu-
midity should
be at least 30 %.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m 30 A/m Power frequen-
cy magnetic
elds should be
at levels
characteristic of
a typical
location in a
typical
commercial or
hospital
environment.