Component Activity Maintenance interval
Installation location, appli‐
ance
Decontamination
The appliance is suitable for room decontamina‐
tion with hydrogen peroxide H
2
O
2
.
In this context, the values of the following appli‐
cation specifications must not be exceeded:
Concentration: maximum 250 ppm.
Humidity: maximum 85%.
Maximum permissible room temperature as per
climate class. (see1.4 Range of appliance use)
Recommendation: as
required/specified at instal‐
lation location, maximum 2x
per year
Fan air slots Cleaning Yearly
Alarm relay
Power cable Check for damage. Yearly
or
if there is a change of loca‐
tion
Tilt angle Check for proper function and tight fit. Yearly
or
if there is a change of loca‐
tion
Adjusting feet/castors Check for proper function, damage, and tight fit.
Replace damaged parts. (see 9.3 Customer
Service)
Yearly
Defrost water collection tray Check the fill level of the defrost water collection
tray and empty if necessary.
Regular visual inspection
8.2 Defrosting the appliance
WARNING
Appliance incorrectly defrosted!
Injuries and damage.
u
Do not use mechanical devices or other means to
accelerate the defrosting process, other than those
recommended by the manufacturer.
u
Do not use any electrical heating or steam cleaning
equipment, naked flames or defrosting sprays.
u
Do not remove ice with sharp objects.
To maintain the functionality of the appliance, we
recommend defrosting it when there is a high buildup
of ice. Icing in the interior is determined by the location
(e.g. high humidity), settings on the appliance and
incorrect user behavior.
If automatic defrost is not sufficient, defrost the appli‐
ance manually:
u
Switch off appliance. (see 6.2.2 Switching appliance
on and off )
u
Disconnect power plug.
u
Store refrigerated items elsewhere.
u
Leave the appliance door open during defrosting
process.
u
Absorb any remaining defrost water with a cloth and
clean the appliance.
8.3 Cleaning the appliance
8.3.1 Preparing
WARNING
Short-circuit hazard due to live parts!
Electric shock or damage to the electronics.
u
Switch off the appliance.
u
Pull out mains plug or switch off fuse.
WARNING
Danger of fire
u
Do not damage the refrigerant circuit.
u
Empty the appliance.
u
Pull out the power plug.
u
Observe the notes on appliance transport.
(see 4.3 Transporting the appliance)
Maintenance