30 OBN-13-BA-e-2511
11.4 Phase contrast units
Basically, there are two different ways to enable the use of phase contrast methods. A
simple and a more extensive method.
1. Simple phase contrast unit
This consists of a simple PH condenser, a PH objective with a specific magnification
(10x, 20x, 40x or 100x), a PH slider, which is adapted to the lens being used, a
centring telescope and two green filters.
To use this, you need to replace the standard condenser of the microscope with
the PH condenser. One of the objectives in the nosepiece is also replaced with the
PH objective and this is positioned in the beam path.