Installation and set-up / operation / maintenance
After mounting, wiring and setting check whether the unit operates
correctly.
Cleaning of the filter cover
If viscous and residues producing
media clog the filter cover of the
sensor (and thus reduce the mea-
suring accuracy slightly), you can
clean it.
Unscrew the filter cover (B) (use a
pair of pliers with plastic-covered
jaws for this). Clean the cover
thoroughly.
The vent (A) should only be clea-
ned by skilled personnel and
with utmost care.
Possible medium residues must not be compressed and pressed
into the vent. This could clog the filter system and reduce the
measuring accuracy of the sensor.
Screw the filter cover again tightly.
The sensor is sufficiently protected against harsh ambient conditions
(protection IP 67). The protection rating can be increased by a special
accessory (order no. E30043).