PG17xx Electronic manometer
10
Zero point behaviour of the analogue output and communication via IO-Link:
5
4
3
1
0
0
5
-
+
+
2
6
6
Fig.2: Analogue output and communication via IO-Link
1: Process value
2: Sensor output signal (green)
3: Ideal measurement signal (red)
4: System pressure
5:
Zero point stabilisation Ò Data sheet
6:
0.5 x zero point stabilisation Ò data sheet
The first process value that can be evaluated via the analogue output and the IO-Link communication
is indicated by the half zero point stabilisation (6).
In the further range of the zero point stabilisation (5), the output signal of the sensor runs at a larger
gradient angle.
The zero point stabilisation is specified in % of the span Ò Data sheet.