7 Operating and display elements
4
2
0
6
8
10
psi
psi
Output
Status
TEMPbar
60
40
20
0
80
100
120
140
bar
Enter
-1
1
2
3
4
8
6
7
5
Fig.9: Operating and display elements (example device type with a measuring range of -1...10 bar; display layout L1)
1: Analogue display of the current system pressure in bar (outer scale) and psi (inner scale)
2: LED ring with 72 segments for colour representation of the analogue display.
A white LED segment extending the pointer indicates the current process value.
3 colour ranges can be configured (Ò Colour scheme of LED ring/43). Example:
Upper colour range: 3.6...6 bar, green
Middle colour range: 2.6...7 bar, yellow
Lower colour range: -1...8.8 bar, red
The display layout of the LED ring can be adjusted (Ò Display layout/42). Depending on the setting: Lag indicator
function for minimum and maximum pressure values.
3: LED pointer: White LED, not visible when the unit is not powered. Indicates the current pressure value between 0 and the
final value of the measuring range in 72 LED steps. The LED segment next to the pointer (pointer extension) lights white.
4: Switching status LED for OUT1: lights yellow if output 1 is active.
5: Digital display (4-digital alphanumeric display):
Indicates the current value. Depending on selection under [SELd]: pressure or temperature.
Indicates the parameters and parameter values.
6: LED provides complementary information about the digital display:
If [SELd] = [PRES], bar (mbar) or psi lights, depending on what is set under [uni.P].
If [SELd] = [TEMP], TEMP lights. The unit set under [uni.T] is displayed.
7: Status LED: The LED lights depending on the device status:
green in normal operation
red in the event of a warning (Ò Warning messages/52) or error (Ò Error messages/52)
8:
Optical operating keys (device keys) for changing the display and for parameter setting (Ò Parameter setting via the unit
keys/29). The device keys can be locked (Ò Lock / unlock/40).
The term “device keys” refers to the touch buttons [Enter], [▲] and [▼].
The touch button [Enter] is indicated by the [●] icon below.
The sensor fields of these optical keys are operated simply by touching the viewing glass. The
touch button must be completely covered to be activated. Slow covering (e.g. liquid flows over
the display) does not activate the touch button.
If the device measures a high internal temperature, the display brightness is automatically
adjusted:
Internal temperature >80°C: The brightness is reduced to 25%.
Internal temperature >90°C: The display is switched off. Only the status LED remains active.
To activate the display, press any key for 30 seconds.