10.4.10.1 Parameter setting via unit keys: lock / unlock
Locking:
u Make sure that the unit is in the normal operating mode.
u Press [▲] + [▼] simultaneously for 10s until [Loc] is displayed.
w The device is locked for parameter setting via the device keys. When trying to change a parameter
value, the symbol [Loc] appears in the digital display.
The locking can only be removed via the device keys. Changing the parameter setting is still
possible via the IO-Link interface.
Unlocking:
u Make sure that the unit is in the normal operating mode.
u Press [▲] + [▼] simultaneously for 10s until [uLoc] is displayed.
w The locking of the device keys is removed.
10.4.11 Reset the device
The device can be reset in two ways.
With both reset applications, the operating hours since the first set-up are not reset.
[APPL] = application reset
The following is reset to the factory setting:
All changed application-specific parameters
If IO-Link data storage is activated, this triggers a parameter update in the master. This writes
the parameters configured in the master to the device again. An application reset may therefore
be ineffective.
[BtB] = Back to Box
The following is reset to the factory setting:
All changed application-specific parameters
All writeable unit identification parameters such as [Application Specific Tag], [Function Tag] or
[Location Tag].
Diagnostic parameters, status parameters, events.
Number of switching cycles
Minimum and maximum memory value for pressure
After the Back to Box reset, the sensor suspends communication and measurement operation
until the voltage is interrupted. The IO-Link data storage is not triggered.
10.4.11.1 Parameter setting via unit keys: reset the unit
u Select [EF] > [rES].
u Select [APPL] or [BtB] and set [Yes].
Only if [BtB] is selected:
u Disconnect and reconnect the voltage supply.
w The device carries out a reboot.