GW Instek GPP-1323 handleiding
Handleiding
Je bekijkt pagina 65 van 157

OPERATION
65
Output
It sets if power output will be activated for the select
step.
Output
ON,OFF
Time
It sets time duration of execution for the select step.
Time
0.1 s to 999.99 s
Voltage
It sets output voltage of CV mode for the select step.
Voltage
0 V to 105 % rated voltage
Current
It sets output limit current of CC mode for the select
step.
Current
0 V to 105 % rated current
OVP Level
It sets over voltage protection setting for the select step.
OVP
Level
5 % to 110 % rated voltage
OCP Level
It sets over current protection setting for the select step.
OCP
Level
5 % to 110 % rated current
Bleeder
It enables or disables discharge loop control for the
select step.
Bleeder
None, ON, OFF
V/I Slew
Rate
It sets High Speed Priority and Slew Rate Priority of CV
and CC modes for the select step.
V/I Slew Rate
CVHS
It utilizes the fastest slew rate of CV mode.
Refer to page 48 for more details.
CCHS
It utilizes the fastest slew rate of CC mode.
Refer to page 51 for more details.
CVLS
It utilizes the user-configured slew rate of
CV mode. When this option is selected, go
to configure the R_V slew Rate (rising) and
F_V slew rate (falling) settings,
respectively. Refer to page 48 for more
details.
CCLS
It utilizes the user-configured slew rate of
CC mode. When this option is selected, go
to configure the R_C slew Rate (rising) and
F_C slew rate (falling) settings,
respectively. Refer to page 51 for more
details.
Buzzer
It enables or disables buzzer sound for the select step.
Buzzer
ON, OFF
Bekijk gratis de handleiding van GW Instek GPP-1323, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
| Merk | GW Instek |
| Model | GPP-1323 |
| Categorie | Niet gecategoriseerd |
| Taal | Nederlands |
| Grootte | 14224 MB |