GPP-1323/GPP-1205 User Manual
22
settings (page 46). For details on how to clear an alarm or to set the
protection modes, please see pages 48.
OVP
Over voltage protection (OVP) prevents a high
voltage from damaging the load. This alarm can be set
by the user.
OCP
Over current protection prevents high current from
damaging the load. This alarm can be set by the user.
OTP
Over temperature protection protect the instrument
from overheating.
Considerations
The following situations should be taken into consideration when
using the power supply.
Inrush
current
When the power supply switch is first turned on, an
inrush current is generated. Ensure there is enough
power available for the power supply when first
turned on, especially if a number of units are turned
on at the same time.
Pulsed or
Peaked loads
When the load has current peaks or is pulsed, it is
possible for the maximum current to exceed the mean
current value. The GPP-1000 power supply ammeter
only indicates mean current values, which means for
pulsed current loads, the actual current can exceed
the indicated value. For pulsed loads, the current
limit must be increased, or a power supply with a
greater capacity must be chosen. As shown below, a
pulsed load may exceed the current limit and the
indicated current on the power supply ammeter.

Bekijk gratis de handleiding van GW Instek GPP-1323, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.

Productinformatie

MerkGW Instek
ModelGPP-1323
CategorieNiet gecategoriseerd
TaalNederlands
Grootte14224 MB