Steps
1. Ensure the power is off on all power supplies.
2. Choose the master and the slave unit(s).
3. Connect the USB cables for the master and slave units
as shown above.
4. Connect the master and slave unit in parallel as
shown above.
5. If using remote sense, connect the master and slave
sense cables as shown above.
Note
Ensure the load cables have sufficient current capacity.
The load wires and remote sense wires should use
twisted-paired wiring of the shortest possible length.
Parallel Operation
Master-
Slave
Configurati
on
Before using the power supplies in parallel, the master
and slave units need to be configured.
Steps
1. Configure the OVP and OCP settings for the master
unit.
2. Enter the Power ON Configuration settings and set
the Tracking configuration for the master and each
connected slave.
3. If using voltage sense, disable local sense in the
Power ON Configuration settings.
4. Cycle the power on the units (reset the power).
Note
Set Tracking to Local to return the units to local
(independent) operation.
Only the Master OVP and OCP settings are used for
protection. Slave protection levels are disregarded.
OTP works independently for each unit.
Master-
Slave
Operation
Only operate the power supplies in parallel if the units
are configured correctly.