GW Instek GPP-1323 handleiding

157 pagina's
PDF beschikbaar

Handleiding

Je bekijkt pagina 113 van 157
DIGITAL CONTROL
113
External Trigger In / Out
Backgroun
d
The trigger input can be configured to perform an
action such as toggling the output on/off, load a
memory setting or apply a voltage/current setting
when a trigger is received.
The trigger output can be configured to be active when
the output is turned on/off, a voltage/current setting is
changed or when a memory setting has been recalled.
The trigger output pulse width can also be configured.
See page 94 for details on the trigger input and trigger
output configuration settings.
Pinout
Name and Pin
Description
TRIG OUT
Trigger output: approx. 3.3 V
Pulse width: approx. 4 ms,
Output impedance: approx.
50 Ω
It outputs a pulse when
power output, V/I set
operation or memory recall is
executed.
TRIG IN
A high or low level TTL
signal is applied for 100 us or
longer.
It receives a pulse to perform
actions like power output,
V/I set operation or memory
recall.
External Operation and Status Monitoring
Overview
The Control I/O Connector is a 10-pin connector that
can be used with the plug for wiring connection. The
pins used determine what remote control mode is used.