Load Cable Connection
Standard
accessories
(GTL-104A,
GTL-105A)
1. Turn the terminal
counterclockwise and loosen
the screw.
2. Insert the cable terminal.
3. Turn the terminal clockwise
and tighten the screw.
Banana plug
Insert the plug into the socket.
Wire type
When using load cables other than the attached,
make sure they have enough current capacity for
minimizing cable loss and load line impedance.
Voltage drop across a wire should not excess 0.5 V.
The following list is the wire current rating at 450
A/cm
2
.
Wire size (AWG)
Maximum current (A)
20
2.5
18
4
16
6
14
10
12
16