Gossen Metrawatt LABKON P500 handleiding
Handleiding
Je bekijkt pagina 57 van 64

LABKON Series
_________________________________________________________________________________________________
GMC-I Messtechnik GmbH 57
Stability, ±(% of output + offset)
Following a 30-minute warm-up, the change occurs in output within 8 hours under constant load, line, and ambient
temperature.
LABKON P500
35 V / 14.5 A
LABKON P500
80 V / 6.5 A
LABKON P500
120 V / 4.2 A
LABKON P800
35 V / 22.5 A
LABKON P800
80 V / 10 A
LABKON P800
120 V / 6.5 A
Voltage
0.02 % + 2 mV
0.02 % + 3 mV
0.02 % + 4 mV
0.02 % + 2 mV
0.02 % + 3 mV
0.02 % + 4 mV
Current
0.02 % + 6 mA
0.02 % + 3 mA
0.02 % +2 mA
0.2 % + 6 mA
0.3 % + 3 mA
0.1 % +2 mA
Voltage Programming Speed
Maximum time required for output voltage to settle within 1% of its total excursion (for resistive load).
Command processing time is excluded.
LABKON P500
35 V / 14.5 A
LABKON P500
80 V / 6.5 A
LABKON P500
120 V / 4.2 A
LABKON P800
35 V / 22.5 A
LABKON P800
80 V / 10 A
LABKON P800
120 V / 6.5 A
Full load Up
50 msec
50 msec
60 msec
50 msec
50 msec
60 msec
Full load
Down
50 msec
50 msec
60 msec
50 msec
50 msec
60 msec
No load Up
50 msec
50 msec
60 msec
50 msec
50 msec
60 msec
No load
Down
200 msec
300 msec
300 msec
200 msec
300 msec
300 msec
Calibration Points
Model
Voltage
Current
K148
0.5 V
22 V
34.5 V
0.5 A
6 A
9 A
K149
0.5 V
44 V
75.5 V
0.5 A
3 A
6 A
K150
0.5 V
60 V
105 V
0.5 A
2 A
4 A
K158
0.5 V
22 V
34.5 V
0.5 A
6 A
9 A
K159
0.5 V
44 V
75.5 V
0.5 A
3 A
6 A
K160
0.5 V
60 V
105 V
0.5 A
2 A
4 A