2.03.2 Hood Power Connection
Installation
Free Standing Ventless Hood
FSH-2 & FSH-2A-99
Model FSH‐2: See Figures 2.03.1‐A ... 2.03.1‐B ... 2.03.1‐C
1. Install appropriate circuit breakers in main electrical panel for hood power. See Section 2.03. for required circuit
breakers. Breaker size given is for the hood only.
2. Remove upper hood back cover & inside electrical box cover. Route appropriate size conduit from panel and attach
with proper connectors at conduit cutout in bracket. Pull proper size wire from electrical panel and attach to
terminal blocks in hood service box. Allow sufficient length of wire & conduit to provide slack needed to
properly access unit for cleaning & service.
3. For hood with optional Appliance Receptacle installed, simply plug power cord from appliance into receptacle
inside hood base, located on back wall. Maximum appliance load = 50A. It may be necessary to change, or install,
the appliance power plug (not provided, customer supplied).
4. For hood without appliance receptacle installed, attach power cord, or wiring, from appliance to terminal blocks
[A1,A2,A3,G] and route through cutout using appropriate strain relief connector or conduit. Maximum appliance
load = 50A.
5. Reinstall inside service cover & hood back panel.
Model FSH‐2A‐99: See Figure 2.03.1‐D
1. Install appropriate circuit breakers in main electrical panel for hood power. See Section 2.03. for required circuit
breakers. Breaker size given is for the hood only.
2. Most FSH‐2A‐99 models have an interlocked power receptacle for the appliance, located on the left hood base
leg, or on the rear panel. When installing a hood with the appliance power receptacle, the breaker size shall be
determined by the electrical load of the appliance being placed under the Hood (Max. load = 50A).
3. Remove the upper rear panel to access terminal blocks. Remove one of the large hole plugs and install an
appropriate conduit connector. Route appropriately sized flexible conduit and wire from the electrical panel
supplying hood power. Allow sufficient length of wire & conduit to provide slack needed to access unit for
cleaning & service.
4. For hood with optional Appliance Receptacle installed, simply plug power cord from appliance into receptacle
inside hood base, located on back wall. Maximum appliance load = 50A. It may be necessary to change, or install,
the appliance power plug (not provided, customer supplied).
5. For hood without appliance receptacle installed, attach power cord, or wiring, from appliance to terminal blocks
[A1,A2,A3,G] and route through cutout using appropriate strain relief connector, or conduit. Maximum appliance
load = 50A.
6. Reinstall inside service cover & hood back panel.
IMPORTANT!
Electrical installation materials (breakers, conduit & fittings, wire, etc.) and labor shall be provided by the customer.
Work should be performed by a qualified professional electrical contactor.
Installation must comply with all local code requirements. Giles assumes no responsibility with respect to code
compliance regarding installation & use of this equipment.