Measurement Menu
Table 7 lists the options available in the Measurement menu.
Level/Span
Level and Span are values within the total range of temperature. Level is the temperature level
to view within the total range of temperatures. Span is the span of temperatures to view within
the total range of temperatures. See Table 8.
Table 7. Measurement Menu
Option Menu Option Description
Level/Span
Auto
Sets the Level/Span to adjust automatically or
manually.
Manual
Set Level/Span
With Level/Span set to Manual, changes the
Level/Span. See Level/Span.
Emissivity
Custom Value
Sets a custom emissivity value when a value
from the standard emissivity table is not
appropriate for the measurement. See Emissivity
Adjustment.
<options>
Select an emissivity value from a list of common
materials. Use the scroll bar to see all of the
options. See Emissivity Adjustment.
Background <options>
Changes the background temperature to
compensate for reflected background
temperature.
Very hot objects or very cold objects can affect
the apparent temperature and measurement
accuracy of the target, especially when surface
emissivity is low. Adjust the reflected background
temperature to improve the accuracy of the
measurement.
Transmissivity <options>
Changes the values used to calculate the
temperature based on the transmission
percentage of the infrared-transparent window
(IR window). See Transmissivity.