Guidance and manufacturer’s declaration-electromagnetic immunity –
for all ME EQUIPMENT AND ME SYSTEMS
Guidance and manufacturer’s declaration-electromagnetic immunity
The 1929LM Touchmonitor for Healthcare Applications is intended for use in the electromagnetic
environment specified below. The customer, or the user of the 1929LM Touchmonitor for Healthcare
Applications,
should assure that it is used in such an environment.
Immunity Test
IEC
60601
test level
Compl
iance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC
61000-4-2
± 6
kV contact
± 8
kV air
± 6 kV
contact
± 8
kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
tr
ansient/burst
IEC
61000-4-4
± 2 kV for power
supply lines
± 1
kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output Lines
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC
61000-4-5
±
1 kV line(s) to
line(s)
±
2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruption
and
voltage
variations
on power supply
input lines
IEC
61000-4-11
<5%
U
T
(>95%
dip in U
T
)
for 0.5 cycle
40%
U
T
(60%
dip in U
T
)
for 5 cycles
70%
U
T
(30%
dip in U
T
)
for 25 cycles
<5%
U
T
(>95%
dip in U
T
)
for 250 cycles
<5%
U
T
(>95%
dip in U
T
)
for 0.5 cycle
40%
U
T
(60
% dip in U
T
)
for 5 cycles
70%
U
T
(30%
dip in U
T
)
for 25 cycles
<5%
U
T
(>95%
dip in U
T
)
for 250 cycles
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the 1929LM Touchmonitor for
Healthcare Applications requires continued
operation during power mains interruptions,
It is recommended that the 1929LM
Touchmonitor for Healthcare Applications be
powered from an uninterruptible power
supply or a battery.
Power Frequency
(50/60
Hz)
Magnetic
Field
IEC 61000-4-8
3
A/m 3 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Bekijk gratis de handleiding van ELO 1929LM, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
| Merk | ELO |
| Model | 1929LM |
| Categorie | Monitor |
| Taal | Nederlands |
| Grootte | 3914 MB |
Caratteristiche Prodotto
| Kleur van het product | Wit |
| Gewicht verpakking | 10000 g |
| Breedte verpakking | 530 mm |
| Diepte verpakking | 265 mm |
| Hoogte verpakking | 480 mm |