07/2023 ba80003def09_EBI25 27
English
Français
hat to do, if
… the display is blank
Maybe the logger or the display is
switched off. Press the start button.
If the display remains blank after
pressing the start button, the battery
has to be replaced.
If the display is activated after press-
ing the start button, switch on the log-
ger or display.
… the logger does not communicate
with the interface
Check if the logger is still within the
radio range. If necessary, reduce the
distance between logger and inter-
face.
Check if the logger is registered on
the interface. If necessary, register
the logger on the interface as de-
scribed in section “Register logger”.
Please run the “Signal test”
Que faire, quand…
… il n’y a pas d’affichage sur le dis-
play
L’enregistreur ou le display pourrait
être désactivé. Appuyez sur le bou-
ton de démarrage.
S’il n’y a pas d’affichage sur le dis-
play après avoir appuyé sur le bouton
de démarrage, vous devez changer
la pile
S’il y a un affichage sur le display,
veuillez réactiver l’enregistreur ou
bien le display
… l’enregistreur ne communique
pas avec l‘interface
Vérifiez si les enregistreurs se
trouvent encore dans la plage de
réception de l’interface. Si
nécessaire veuillez réduire la
distance entre l’enregistreur et
l’interface.
Vérifiez si l’enregistreur est connec
à l’interface. Veuillez faire la connex-
ion comme décrite chez «Connecter
l’enregistreur».
Veuillez faire un «Test de signal»