4.6.2. Manual Arc Calibra�on
Cut and clamp fiber according to splice procedure, and then
press the key . The calibra�on program starts to run and test
results are given. According to the test results, manually set the
splicing parameters of Arc data, and repeat the above steps
un�l the test results meet the requirements.
Parameter modifica�on requirements: the default value is
between 8-12, every 5 to modify 1 Arc data value. When the
calibra�on result is larger than 12, we should reduce Arc data
value. Or users can choose Auto Arc Calibra�on directly. Shown
in below Figure 4.24.
Figure 4.24