DWL-5x000XY DATA SHEET
Byte [n] Command Descriptions
Byte [0] 0x61 To Computer / From Sensor
Byte [1] 0x13 Single Axis Position / Vibro Mode
Byte [2] 0x00
Vibro Value
Byte [3] 0x03
Byte [4] 0xD0
Byte [5] 0x90
Byte [6] 0xCC
Fixed Dummy
Byte [7] 0xCC
Byte [8] 0xCC
Byte [9] 0xCC
Byte [10] 0xAA CRC16 High
Byte [11] 0x3A CRC16 Low
Following equation shows the conversion of angle from the received data:
For DWL5800XY
Vibrometer g = (((Byte [5] << 24) + (Byte [4] << 16) + (Byte [3] << 8) + Byte [2]) - 250000) /
100000
For DWL5500XY
Vibrometer g = (((Byte [5] << 24) + (Byte [4] << 16) + (Byte [3] << 8) + Byte [2]) - 250000) / 100000
For DWL5000XY
Vibrometer g = (((Byte [5] << 24) + (Byte [4] << 16) + (Byte [3] << 8) + Byte [2]) - 25000) / 10000
6.5.4. Calibration Mode
To set the tilt sensor module into Calibration Mode:
<< (To sensor) : 0x06, 0x01, 0x0B, 0xA0, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00
<< (To sensor) : 0x06, 0x01, 0x0B, 0xB0, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00
Byte [n] Command Descriptions
Byte [0] 0x06 From Computer
Byte [1] 0x01 To Sensor
Byte [2] 0x0B Calibration Mode
Byte [3] 0xA0 0xA0 : Acknowledgement
0xB0 : Start current calibration step
Byte [4]~[11] 0x00 Not Applicable
Data received from tilt sensor module:
>> (From sensor) : 0x61, 0x1B, 0x00, 0x00, 0x13, 0x01, 0x88, 0xDD, 0xDD, 0xDD, 0xAA, 0x3A
Byte [n] Command Descriptions
Byte [0] 0x61 To Computer / From Sensor
Byte [1] 0x1B Single Axis Position / Calibration Mode
Byte [2] 0x00 Calibration Feedback
0x17 : Idle /Current calibration step nished
0x0A : Counting
Byte [3] 0x00 Calibration Count
0x01 - 0x1E : Counting Up
Byte [4] 0x13 Not Applicable
Byte [5] 0x01 Calibration Step
0x01: Calibration Step 1 ready
0x02: Calibration Step 2 ready
0x03: Calibration Step 3 ready
0x04: Calibration Step 4 ready
0x05: Calibration Step 5 ready
0x06: Calibration Step 6 ready
0x07: Calibration Step 7 ready
0x08: Calibration Step 8 ready
Byte [6] 0x88 Not Applicable
Byte [7]~[9] 0xDD Fixed Dummy
Byte [10] 0xAA CRC16 High
Byte [11] 0x3A CRC16 Low
It is advised to perform calibration of the tilt sensor module in a controlled temperature set-
ting.
* Please refer to 6.5.4.1 for detailed calibration steps
6.5.4.1. Detailed Calibration Steps :
1. Please refer to Appendix 1 for proper instrument setup for calibration. Send the follow-
ing command from computer to respective sensor to make sure the sensor is ready for
calibration mode:
<< (To sensor): 0x06, 0x01, 0x0B, 0xA0, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00
Byte [n] Command Descriptions
Byte [0] 0x06 From Computer
Byte [1] 0x01 To Sensor
Byte [2] 0x0B Calibration Mode
Byte [3] 0xA0 Acknowledgement to sensor module
Byte [4]~[11] 0x00 Not Applicable