Chauvin Arnoux L461 handleiding
Handleiding
Je bekijkt pagina 19 van 48

19
4) Connecting to the DataViewSync
TM
(IRD server)
■ In Data Logger Transfer, change the connection by clicking on
then on DataViewSync
TM
(IRD server).
■ Enter the DataViewSync
TM
(IRD server) address (the same one chosen during setup), the instrument serial number and the
password you set in the previous step.
■ Click Next to conrm.
3.2.3. RATED PRIMARY CURRENT (L411, L412)
Press the ▼ key to go to the next screen.
Figure 15
For the L412:
■ Connect the current sensor(s).
■ The current sensor is automatically detected by the instrument.
■ If two current sensors are connected, they must be identical.
For AmpFlex
®
or MiniFlex
sensors, press the key to select 300 or 3000 A.
For the other sensors, conguration is done via Data Logger Transfer.
The rated currents of the current sensors are as follows:
Sensor Rated current Choice of gain Number of turns
C193 clamp 1000 A
AmpFlex
®
A193
MiniFlex MA194 300 or 3,000 A
1, 2 or 3
to be congured in Data Logger
Transfer
MN93A clamp 5 A calibre 5 to 25,000 A
to be congured in Data Logger
Transfer
MN93A clamp 100 A calibre 100 A
MN93 clamp 200 A
MINI 94 clamp 200 A
Table 4
3.2.4. AGGREGATION PERIOD
Press the ▼ key to go to the next screen.
Figure 16
Number of turns the toroid makes
around the conductor.
To change the aggregation period, press
the
key: 1, 2, 3, 4, 5, 6, 10, 12, 15,
20, 30 or 60 minutes.
Bekijk gratis de handleiding van Chauvin Arnoux L461, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
Merk | Chauvin Arnoux |
Model | L461 |
Categorie | Niet gecategoriseerd |
Taal | Nederlands |
Grootte | 4956 MB |