2.3. FAULT SIMULATION
Once the functional test has been carried out successfully, perform a fault simulation
2.3.1. SHORT CIRCUIT BETWEEN CP AND PE
Pressing the CP-PE short button simulates a short circuit between the CP signal and the PE protection
conductor for 3 seconds.
The charging station must lock for 30 seconds. The terminal indicates that it is locked. The charging
process underway is stopped. After 30 seconds, the terminal reports that there is no more fault and unlocks.
This test must be conducted on each charging mode: A, B, C and D .
2.3.2. SHORT CIRCUIT BY DIODE BETWEEN CP AND PE
Pressing the DIODE short button simulates a short circuit by diode between the CP signal and the PE
protection conductor. This fault duration is the same as the length of the button press.
The charging terminal must deactivate within 30 seconds. The terminal indicates that it is locked. The
charging process underway is stopped. After 30 seconds, the terminal reports that there is no more fault
and unlocks.
This test must be conducted on each charging mode: A, B, C and D .
2.3.3. PE OPEN
Pressing the PE open button simulates disconnection of the PE protection conductor.
The charging terminal must deactivate within 100 ms. The terminal is locked and only an EVCI approved
installer can reactivate it.
If any of these 3 faults fails to lock the terminal, the tests must be stopped and the terminal must
be repaired.