43
Problem Possible cause Solution
Fastener cannot be
opened
Incorrect handling Open the fastener as described in
chapter “7.3 Opening/closing the
wristband”.
Wristband size can-
not be adjusted
Incorrect handling Adjust the wristband size as
described in chapter “7.4 Adjusting
the wristband size”.
ON/OFF/emergency
button
3
does not
work
Battery is empty Charge the battery as described in
chapter “7.2 Charging the battery”.
Device is switched o Switch on the device as described
in chapter “7.5 Switching the emer-
gency wristband on and o”
ON/OFF/emergency button
3
is faulty
Contact Customer Services
LEDs
2
are not
working
Battery is empty Charge the battery as described in
chapter “7.2 Charging the battery”.
Device is switched o Switch on the device as described
in chapter “7.5 Switching the emer-
gency wristband on and o”
The LEDs
2
are faulty
Contact Customer Services
Emergency notifica-
tion is not success-
fully transferred (no
green running light)
No network connection Check the network strength at
this location as described in
chapter “7.10 Checking the network
strength”.
Automatic emergen-
cy notification is not
triggered following
a fall
Battery is empty Charge the battery as described in
chapter “7.2 Charging the battery”.
Device is switched o Switch on the device as described
in chapter “7.5 Switching the emer-
gency wristband on and o”
No fall was detected See chapter 7.7 for further infor-
mation about the fall detection
function. The impact may have
been too weak.