3.1 Scheduled maintenance
About
The MUIP-2213 does not require any scheduled maintenance or calibration activities. In case of
inconsistencies, please contact Barco Healthcare.
3.2 Cleaning
WARNING: Unplug the power cable from the mains power input before cleaning the display.
CAUTION: Take care not to damage or scratch the front glass or LCD. Be careful with rings or other
jewelry and do not apply excessive pressure on the front glass or LCD.
CAUTION: Do not apply or spray liquid directly to the display as excess liquid may cause damage to
internal electronics. Instead, apply the liquid to a cleaning cloth.
To clean the display
Clean the display using a sponge, cleaning cloth or soft tissue, lightly moistened with a recognized cleaning
product for medical equipment. Read and follow all label instructions on the cleaning product. In case of doubt
about a certain cleaning product, use plain water.
Possible cleaning solutions:
Isopropryl alcohol 90%
Flux
Ethanol 70%
Chlorine 250 ppm in 1-liter distilled water
Wurth, TFT-Reininger
Soap with lukewarm water
1.6% ammonia
Sodium hypochlorite 10%
Peroxide hydrogen10%
Tanet interior
Do not use following products:
Alcohol/solvents at higher concentration > 70%
Strong alkalis lye, strong solvents
Acid
Detergents with fluoride
Detergents with ammonia at higher concentration > 1.6%
Detergents with abrasives
Steel wool
Sponge with abrasives
Steel blades
Cloth with steel thread
Maintenance