movement. Decentration and excessive lid sensation accompanied
by limited movement often indicates the lens is too at for the given
eye.
If the criteria for a well-tted lens cannot be achieved, do not dispense.
C. Final Lens Power Determination
After the characteristics of a well-tted lens have been satised, conduct
a spherical over-refraction to determine the proper lens power to be
dispensed.
Example:
Trial lens: -4.50D
Over-refraction: -0.25D
Final Lens Power: -4.75D
Use a fresh, new pair of lenses for each trial tting. Do not attempt to disinfect
and re-use trial lenses.
FITTING GUIDELINES (Toric Lenses)
The geometry of a FOCUS
®
DAILIES
®
Toric and DAILIES
®
AquaComfort
Plus
®
Toric lens is a double thin zone design. The back surface tri-curve
toric design is available in one base curve and ts a wide variety of eyes,
reducing tting time and inventory requirements. The FOCUS
®
DAILIES
®
Toric and DAILIES
®
AquaComfort Plus
®
Toric lens design has a constant
thickness difference between the vertical, thin zones and a horizontal thicker
zone, resulting in consistent and excellent stabilization over the power range.
To aid the tting process, FOCUS
®
DAILIES
®
Toric and DAILIES
®
AquaComfort Plus
®
Toric lenses feature scribe lines on the front lens
surface to enable assessment of the lens orientation. These lines are at
3 and 9 o’clock positions approximately 1.0 mm in from the lens edge.
The vertical line of the "K" on the "OK" inversion mark coincides with the
90 degree meridian of the lens and thus might also be used for the
assessment of the lens orientation. The lens orientation ndings are then
used for calculation of axis compensations.
1. Patient Selection
The patient characteristics necessary to achieve success with FOCUS
®
DAILIES
®
Toric and DAILIES
®
AquaComfort Plus
®
Toric lenses are similar
to those for spherical lenses. A thorough pre-tting examination should be
conducted to ensure the astigmatic patient is a suitable candidate for soft
contact lens wear.
The following procedures should be followed when tting FOCUS
®
DAILIES
®
Toric and DAILIES
®
AquaComfort Plus
®
Toric lenses. For additional tips
on tting the monovision patient refer to the section Monovision Fitting
Guidelines.
2. Pre-Fitting Examination
A pre-tting examination is necessary to:
• assessthepatient’smotivation,physicalstateandwillingnessto
comply with instructions regarding hygiene and wear schedule
Bekijk gratis de handleiding van Alcon AquaComfort Plus, stel vragen en lees de antwoorden op veelvoorkomende problemen, of gebruik onze assistent om sneller informatie in de handleiding te vinden of uitleg te krijgen over specifieke functies.
Productinformatie
| Merk | Alcon |
| Model | AquaComfort Plus |
| Categorie | Niet gecategoriseerd |
| Taal | Nederlands |
| Grootte | 4631 MB |
Caratteristiche Prodotto
| Materiaal | Nelfilcon A |
| Aantal per verpakking | 90 stuk(s) |
| Type product | Dagelijks |
| Ultraviolet ( UV) bescherming | Nee |
| Lens diameter | 14 mm |