POC-621-12 Series User Manual 18
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The POC-621-12 series device is intended for use in the electromagnetic environment
specified below. The customer or the user of the model POC-621-12 series device should
ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environ-
ment – guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Interruptions and
voltage varia-
tions on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for input/out-
put lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Voltage Dips:
i) 0% reduction
for 0.5 cycle at
50Hz,
0% reduction
for 1 cycle at 50Hz
Performance
Criterion B
ii) 70% reduction
for 25/30 cycles at
50/60Hz
Performance
Criterion C
Voltage Interrup-
tions:
0% reduction
for 250/300 cycles
at 50/60Hz
Performance
Criterion C
30A/m
±8 kV contact
±15 kV air
±2 kV for
power supply
lines
±1 kV for input/
output lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Voltage Dips:
i) 0% reduc-
tion for 0.5
cycle at 50Hz,
0% reduc-
tion for 1 cycle
at 50Hz
Performance
Criterion B
ii) 70%
reduction for
25/30 cycles at
50/60Hz
Performance
Criterion C
Voltage Inter-
ruptions:
0% reduc-
tion for 250/
300 cycles at
50/60Hz
Performance
Criterion C
30A/m
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Main power quality should be
that of a typical commercial or
hospital environment.
If the user of the model POC-
621-12 requires continued
operation during main power
interruption, it is recom-
mended that the model POC-
621-12 be powered from an
uninterruptible power supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.
NOTE: UT is the AC mains voltage prior to application of the test level.