FUNCTION DETAILS
Temperature Control: After turning on the unit and reaching the delay time (parameter E40), the compressor
starts operating when the cabinet temperature is higher than (set temperature + hysteresis) and will be off when
the cabinet temperature is lower than the set point temperature.
T
o protect the compressor, it cannot be re-started unless the time after the compressor stops is longer
than the delay time (Parameter E41).
D
efrost Control: After running for a defrost interval set time (parameter F2), the unit will enter defrost and
the compressor stops. When the defrost duration time (parameter F1) ends, it will exit defrost status. After
a defrost period ends, there will be 2 minutes of dripping time before the unit can enter refrigeration
status.
• When defrost interval F2 is set to 00, the automatic defrost by turning off compressor will be
cancelled.
• When setting parameter F4 is set to 0, the temperature will be displayed normally during
defrost.
• When setting parameter F4 is set to 1, the cabinet temperature is locked during defrost, and the
last value before defrost is displayed. When defrost ends, normal display will resume after
temperature display 20 minutes delay (or cabinet temperature lower than the set temperature).
The defrost LED flashes during delay.
• When setting parameter F4 is set to 2, the set temperature will be displayed during defrost.
• When defrost ends, normal display will be resumed after 20 minutes delay (or cabinet
temperature lower than the set temperature). The defrost LED flashes during delay.
• When setting parameter F4 is set to 3, [dEF] will be displayed during defrost. When defrost ends,
normal display will be resumed after 20 minutes delay display [dEF] (or cabinet temperature lower
than the set temperature). The defrost LED flashes during delay.
A
larm Control: After turning on for the first time, the unit will need to reach the [C5] delay time setting, then
the high/low temperature alarm function can be triggered (C1, C2, C3). After passing [C5] delay, when the
cabinet temperature is abnormal (for example higher than the temperature alarm [C1] or lower than
temperature alarm [C2]) and duration more than alarm delay time [C6], the unit will enter alarm status,
alarm start. When high temperature alarm alternate display H1 and cabinet air temperature, the compressor
starts to run. When low temperature alarm, will alternate display LO and cabinet temperature, the
compressor stops. When cabinet temperature is higher than the (low temperature alarm value [C2] + alarm
hysteresis [C4]), the low temperature alarm ends. When cabinet temperature is lower than (high temperature
alarm value [C1] –alarm hysteresis [C4], the high temperature alarm ends.
When the product sensor temperature is greater than or equal to the product temperature high temperature
alarm value [C3] and duration more than alarm delay time [C6], unit enters alarm status, and starts the alarm,
alternate display [H2] and cabinet temperature. When the product sensor temperature is less than or equal
to the (high temperature alarm value [C3] – alarm hysteresis [C4]), it will end the high temperature alarm.
W
hen the electric supply is off, flash and display [EEL] and alarm. When setting do1 is set at 00, unit will not
alarm when door is open. When setting do1 is set at 0, once unit reaches the delay time, it will flash and
display [dr alarm], press any button to cancel.
If se
tting [C7] is set to 0, then the units relay does not alarm when power is off.